| ControlChannel. Set/get which input to control from. Valied values: A and B. |
| ControlMode. Configures/queries the control loop mode. 1=manual PID, 2= Zone, 3= Open loop, 4= AutoTune PID, 5= AutoTune PI, 6= AutoTune P. |
| ControlModeInfo. Shows the control loop mode in text. |
| Derivative. Writes/reads D parameter. |
| HeaterRange. Configures/queries the heater range. Valid entries: 0 - 5. Setting the range to 0 turns the heater off. |
| Integral. Writes/reads I parameter. |
| InputCurveA. Writes/reads the curve the channel A uses for temperature conversion. Valid entries: 0 = none, 1 - 20 = standard curves, 21 - 60 = user curves. |
| InputCurveA. Writes/reads the curve the channel B uses for temperature conversion. Valid entries: 0 = none, 1 - 20 = standard curves, 21 - 60 = user curves. |
| OutputChannelA. Kelvin reading for channel A. |
| OutputChannelB. Kelvin reading for channel B. |
| Proportional. Writes/reads P parameter. |
| RampRate. Configures/queries the number of kelvin per minute to ramp the setpoint. |
| RampOffOn. If 0/1 the ramping is off/on. By writting 0/1 turns the ramping off/on. |
| SetPoint. Writes/reads the temperature set point (in Kelvin). |