Properties

pogo_html_docu/FMBOxfDCMEnergy/Properties.html See pogo generated info for more details.

 Beamline. Beamline number. Used for getting the shutter status if the commands startup or shutdown are called.

 BraggDevice. Name of the FMBOxfDCMMotor tango device corresponding to the Bragg Motor.

 BraggOffsetC0. UnitCalibration to be written to the Bragg Motor using the ChangeCrystalAndRecalibrate command for changing to crystal 0.

 BraggOffsetC1. UnitCalibration to be written to the Bragg Motor using the ChangeCrystalAndRecalibrate command for changing to crystal 1.

 BraggOffsetC2. UnitCalibration to be written to the Bragg Motor using the ChangeCrystalAndRecalibrate command for changing to crystal 2.

 Bragg2Device. Name of the FMBOxfDCMMotor tango device corresponding to the 2nd. Bragg Motor in the DCM GKSS monochromator (only used for version 10).

 Bragg2OffsetC0. UnitCalibration to be written to the Bragg Motor in the DCM GKSS monochromator (only used for version 10) using the ChangeCrystalAndRecalibrate command for changing to crystal 0.

 Bragg2OffsetC1. UnitCalibration to be written to the Bragg Motor in the DCM GKSS monochromator (only used for version 10) using the ChangeCrystalAndRecalibrate command for changing to crystal 1.

 Bragg2OffsetC2. UnitCalibration to be written to the Bragg Motor in the DCM GKSS monochromator (only used for version 10) using the ChangeCrystalAndRecalibrate command for changing to crystal 2.

 CrystalCutGap. Value of the gap when the crystal cut is used (crystal 2). Only used with crystal 2 for computing the ExitOffset as a function of the Bragg angle.

 DCrystalSI111. Lattice spacing for SI111 crystal. If value is not set or set to 0 or negative, a default value is taken.

 DCrystalSI311. Lattice spacing for SI311 crystal. If value is not set or set to 0 or negative, a default value is taken.

 FMBOxfDCMCtrlDevice. Name of the FMBOxfDCMCtrl tango device making the connection to the socket. Not used for the petra3 extension beamlines.

 FMBOxfDCMCtrlDevice2. Name of the a second FMBOxfDCMCtrl tango device for the case of having more motors (only used for versions bigger than 9).

 LongitudinalDevice. Name of the FMBOxfDCMMotor tango device corresponding to the longitudinal motor in the DCM GKSS monochromator (only used for version 10).

 LongitudinalOffsetC0. UnitCalibration to be written to the longitudinal Motor in the DCM GKSS monochromator (only used for version 10) using the ChangeCrystalAndRecalibrate command for changing to crystal 0.

 LongitudinalOffsetC1. UnitCalibration to be written to the longitudinal Motor in the DCM GKSS monochromator (only used for version 10) using the ChangeCrystalAndRecalibrate command for changing to crystal 1.

 LongitudinalOffsetC2. UnitCalibration to be written to the longitudinal Motor in the DCM GKSS monochromator (only used for version 10) using the ChangeCrystalAndRecalibrate command for changing to crystal 2.

 ParDevice. Name of the FMBOxfDCMMotor tango device corresponding to the parallel motor. Only for versions smaller than 10.

 ParOffsetC0. UnitCalibration to be written to the Parallel Motor using the ChangeCrystalAndRecalibrate command for changing to crystal 0. Only for versions smaller than 10.

 ParOffsetC1. UnitCalibration to be written to the Parallel Motor using the ChangeCrystalAndRecalibrate command for changing to crystal 1. Only for versions smaller than 10.

 ParOffsetC2. UnitCalibration to be written to the Parallel Motor using the ChangeCrystalAndRecalibrate command for changing to crystal 2. Only for versions smaller than 10.

 PerpDevice. Name of the FMBOxfDCMMotor tango device corresponding to the perpendicular motor. Only for versions smaller than 10.

 PerpOffsetC0. UnitCalibration to be written to the Perpendicular Motor using the ChangeCrystalAndRecalibrate command for changing to crystal 0. Only for versions smaller than 10.

 PerpOffsetC1. UnitCalibration to be written to the Perpendicular Motor using the ChangeCrystalAndRecalibrate command for changing to crystal 1. Only for versions smaller than 10.

 PerpOffsetC2. UnitCalibration to be written to the Perpendicular Motor using the ChangeCrystalAndRecalibrate command for changing to crystal 2. Only for versions smaller than 10.

 SimulationMode. 0 real mode, 1 simulation mode.

 Version. 0 -> old PMAC software, 1 -> new PMAC software, 2 -> new PMAC software with old conversion factors (p10), 10 -> DCM from GKSS, different energy calculation, 11 -> SBM from GKSS, different energy calculation. For the P3 extension beamlines any number lower than 9 can be used.