The following attributes are needed to start the T2 or T3 mode. They are taken from the demo code tttrmode.c provided by PicoQuant. Given are the attribute names and the setter functions.
Manual: The sync divider must be used to keep the effective sync rate at values lessEqual 12.5 MHz. It should only be used with sync sources of stable period. Using a larger divider than strictly necessary does not do great harm but it may result in slightly larger timing jitter. The readings obtained with HH_GetCountRate are internally corrected for the divider setting and deliver the external (undivided) rate. The sync divider should not be changed while a measurement is running.
CFD discriminator level in millivolts, between DISCRMIN, DISCRMAX CFD zero cross level in millivolts, between ZCMIN, ZCMAX
sync timing offset in ps, between CHANOFFSMIN, CHANOFFSMAX
CFD discriminator level in millivolts, between DISCRMIN and DISCRMAX CFD zero cross level in millivolts, between ZCMIN and ZCMAX
channel timing offset in ps, between CHANOFFSMIN and CHANOFFSMAX
In T3 mode, binning and offset are additionally set.
measurement binning code, minimum = 0 (default = smallest, i.e. base resolution) maximum = (MAXBINSTEPS-1) (largest)
histogram time offset in ns, between OFFSETMIN (0, default) and maximum = OFFSETMAX