Command Prime :
Prime the pump. The controller has no command for this, so we do
it in software as follows:
- set the flow rate to value given in property PrimeFlowRate (default 24.0)
- set the upper pressure limit to value given in property PrimePressure (default 350)
- start the pump
- wait until upper pressure limit is reached, i.e. the corresponding fault flag is set
- reset flow rate and upper pressure limit to what they were before
Prime Definition |
Input Argument |
Tango::DEV_VOID |
|
Output Argument |
Tango::DEV_VOID |
|
DisplayLevel |
OPERATOR |
.. |
Inherited |
false |
.. |
Abstract |
false |
.. |
Polling Period |
Not polled |
.. |
|
|
|
Command NOT allowed for |
FAULT OFF |
.. |