MKS651C Class Attributes
Name
Inherited
Abstract
Attr. type
R/W type
Data type
Level
Description
SetPointA false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR Set point A value

unit depends on AbsoluteValues setting:

if true, value is pressure
if false, value is % of full scale range
AbsoluteValues false false Scalar READ_WRITE Tango::DEV_BOOLEAN OPERATOR The controller returns values for presssure settings (pressure,
setpoint, process limit) in % of full range.

If you want the real values calculated set this to true.
SetPointB false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR Set point B value

unit depends on AbsoluteValues setting:

if true, value is pressure
if false, value is % of full scale range
SetPointC false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR Set point C value

unit depends on AbsoluteValues setting:

if true, value is pressure
if false, value is % of full scale range
SetPointD false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR Set point D value

unit depends on AbsoluteValues setting:

if true, value is pressure
if false, value is % of full scale range
SetPointE false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR Set point E value

unit depends on AbsoluteValues setting:

if true, value is pressure
if false, value is % of full scale range
SetPointAnalogue false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR Analogue set point A value

unit depends on AbsoluteValues setting:

if true, value is pressure
if false, value is % of full scale range
SelectedSetPoint false false Scalar READ_WRITE Tango::DEV_USHORT OPERATOR Selected setpoint:

0 = analogue
1 = A
2 = B
3 = C
4 = D
5 = E
Unit false false Scalar READ_WRITE Tango::DEV_USHORT OPERATOR Pressure Unit:

0 = Torr
1 = mTorr
2 = mbar
3 = nbar
4 = kPa
5 = Pa
6 = cmH2O
7 = inH2O

This just changes the unit sign displayed on the device.
Check / set SensorPressureRange value as well.
SensorPressureRange false false Scalar READ_WRITE Tango::DEV_USHORT OPERATOR Sensor pressure range:

0 = 0.1
1 = 0.2
2 = 0.5
3 = 1.0
4 = 2.0
5 = 5.0
6 = 10.0
7 = 50.0
8 = 100.0
9 = 500.0
10 = 1000.0
11 = 5000.0
12 = 10000.0
13 = 1.33
14 = 2.66
15 = 13.33
16 = 133.3
17 = 1333.0
18 = 6666.0
19 = 13332.0
SensorVoltageRange false false Scalar READ_WRITE Tango::DEV_USHORT OPERATOR Sensor voltage range:

0 = 1 V
1 = 5 V
2 = 10 V
ValveState false false Scalar READ_WRITE Tango::DEV_USHORT OPERATOR Valve State:

0 = open
1 = close
2 = stop
SoftStartA false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR softstart rate of set point A, value is % of full speed
SoftStartB false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR softstart rate of set point B, value is % of full speed
SoftStartC false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR softstart rate of set point C, value is % of full speed
SoftStartD false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR softstart rate of set point D, value is % of full speed
SoftStartE false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR softstart rate of set point E, value is % of full speed
SoftStartAnalogue false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR softstart rate of analog set point, value is % of full
speed
SoftStartValveOpen false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR softstart rate of open valve, value is % of full speed
SoftStartValveClose false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR softstart rate of close valve, value is % of full speed
ProcessLimit1Low false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR low threshold process limit 1

unit depends on AbsoluteValues setting:

if true, value is pressure
if false, value is % of full scale range
ProcessLimit1High false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR high threshold process limit 1

unit depends on AbsoluteValues setting:

if true, value is pressure
if false, value is % of full scale range
ProcessLimit2Low false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR low threshold process limit 2

unit depends on AbsoluteValues setting:

if true, value is pressure
if false, value is % of full scale range
ProcessLimit2High false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR high threshold process limit 2

unit depends on AbsoluteValues setting:

if true, value is pressure
if false, value is % of full scale range
AnalogueSetPointRange false false Scalar READ_WRITE Tango::DEV_USHORT OPERATOR Sensor pressure range:

0 = 0.1
1 = 0.2
2 = 0.5
3 = 1.0
4 = 2.0
5 = 5.0
6 = 10.0
7 = 50.0
8 = 100.0
9 = 500.0
10 = 1000.0
11 = 5000.0
12 = 10000.0
13 = 1.33
14 = 2.66
15 = 13.33
16 = 133.3
17 = 1333.0
18 = 6666.0
19 = 13332.0
SetPointAnalogueType false false Scalar READ_WRITE Tango::DEV_USHORT OPERATOR Set point A type where value is:
0 = position
1 = pressure
SetPointBType false false Scalar READ_WRITE Tango::DEV_USHORT OPERATOR Set point A type where value is:
0 = position
1 = pressure
SetPointCType false false Scalar READ_WRITE Tango::DEV_USHORT OPERATOR Set point A type where value is:
0 = position
1 = pressure
SetPointDType false false Scalar READ_WRITE Tango::DEV_USHORT OPERATOR Set point A type where value is:
0 = position
1 = pressure
SetPointEType false false Scalar READ_WRITE Tango::DEV_USHORT OPERATOR Set point A type where value is:
0 = position
1 = pressure
SetPointAType false false Scalar READ_WRITE Tango::DEV_USHORT OPERATOR Set point A type where value is:
0 = position
1 = pressure
ValvePositionOutputRange false false Scalar READ_WRITE Tango::DEV_USHORT OPERATOR Valve position output range:

0 = 5 V
1 = 10 V
DirectReverseControl false false Scalar READ_WRITE Tango::DEV_USHORT OPERATOR Direct / Reverse control:
0 = direct
1 = reverse
SensorType false false Scalar READ_WRITE Tango::DEV_USHORT OPERATOR Sensor type:
0 = absolut
1 = differential
LeadA false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR Lead for Setpoint A
GainA false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR gain of Setpoint A, where value = % gain
LeadB false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR Lead for SetpointB
LeadC false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR Lead for Setpoint C
LeadD false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR Lead for Setpoint D
LeadE false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR Lead for Setpoint E
GainB false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR gain of Setpoint B, where value = % gain
GainC false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR gain of SetpointC, where value = % gain
GainD false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR gain of SetpointD, where value = % gain
GainE false false Scalar READ_WRITE Tango::DEV_DOUBLE OPERATOR gain of Setpoint E, where value = % gain
ControlType false false Scalar READ_WRITE Tango::DEV_BOOLEAN OPERATOR Control type:

0 = self-tuning
1 = PID
ValveOnPowerFail false false Scalar READ_WRITE Tango::DEV_USHORT OPERATOR Behaviour of valve on power failure:

0 = option not activated or installed
1 = valve opens on power failure
2 = valve closes on power failure
ValveType false false Scalar READ_WRITE Tango::DEV_USHORT OPERATOR Sensor type:
0 = absolut
1 = differential
Pressure false false Scalar READ Tango::DEV_DOUBLE OPERATOR
ValvePosition false false Scalar READ Tango::DEV_DOUBLE OPERATOR