OxfordCryostream700 Class Attributes |
Name |
Inherited |
Abstract |
Attr. type |
R/W type |
Data type |
Level |
Description |
GasSetPoint |
false |
false |
Scalar |
READ |
Tango::DEV_FLOAT |
OPERATOR |
Set temperature. |
GasTemp |
false |
false |
Scalar |
READ |
Tango::DEV_FLOAT |
OPERATOR |
Current gas temperature. |
GasError |
false |
false |
Scalar |
READ |
Tango::DEV_FLOAT |
EXPERT |
Difference between current and set temperature. |
RunMode |
false |
false |
Scalar |
READ |
Tango::DEV_STRING |
EXPERT |
RunMode of the cryo controller (StartUp, StartUpFail, StartUpOk, Run, SetUp, ShutdownOk, ShutdownFail). |
Phase |
false |
false |
Scalar |
READ |
Tango::DEV_STRING |
EXPERT |
Program Phase of the cryo controller (Ramp, Cool, Plat, Hold, End, Purge, Soak, Wait). This parameter is meaningless unless RunMode is Run. |
RampRate |
false |
false |
Scalar |
READ |
Tango::DEV_USHORT |
OPERATOR |
Cooling ramp rate for the current program phase. |
TargetTemp |
false |
false |
Scalar |
READ_WRITE |
Tango::DEV_FLOAT |
OPERATOR |
Temperature set point for the current program phase. If a value is written, the device will ramp up with 360K/h or cool down to it. |
Remaining |
false |
false |
Scalar |
READ |
Tango::DEV_USHORT |
EXPERT |
Time remaining in current program phase. |
GasFlow |
false |
false |
Scalar |
READ |
Tango::DEV_FLOAT |
OPERATOR |
Current gas flow. |
GasHeat |
false |
false |
Scalar |
READ |
Tango::DEV_UCHAR |
EXPERT |
Current gas heating. |
LinePressure |
false |
false |
Scalar |
READ |
Tango::DEV_FLOAT |
EXPERT |
Current back pressure. |
Alarm |
false |
false |
Scalar |
READ |
Tango::DEV_STRING |
OPERATOR |
Indicator for most serious alarm conditions. |
RunTime |
false |
false |
Scalar |
READ |
Tango::DEV_STRING |
EXPERT |
Uptime of the pump. |
ControllerNumber |
false |
false |
Scalar |
READ |
Tango::DEV_USHORT |
EXPERT |
Serial number of the controller device. |
SoftwareVersion |
false |
false |
Scalar |
READ |
Tango::DEV_UCHAR |
EXPERT |
Firmware version of the controller device. |
Hardware |
false |
false |
Scalar |
READ |
Tango::DEV_STRING |
EXPERT |
Connected controller and cooler system devices. |
OxfordCryostream700 Class Dynamic Attributes |
Name |
Inherited |
Abstract |
Attr. type |
R/W type |
Data type |
Level |
Description |
ColdheadSerial |
false |
false |
Scalar |
READ |
Tango::DEV_USHORT |
EXPERT |
N-Helix coldhead serial number. |
EvapTemp |
false |
false |
Scalar |
READ |
Tango::DEV_FLOAT |
EXPERT |
Current evaporator temperature. |
SuctTemp |
false |
false |
Scalar |
READ |
Tango::DEV_FLOAT |
EXPERT |
Current SUCT temperature. |
RefrTemp |
false |
false |
Scalar |
READ |
Tango::DEV_FLOAT |
EXPERT |
Current refrigerator temperature. |
ShieldTemp |
false |
false |
Scalar |
READ |
Tango::DEV_DOUBLE |
EXPERT |
Current shield temperature. |
NozzleTemp |
false |
false |
Scalar |
READ |
Tango::DEV_FLOAT |
EXPERT |
Current nozzle temperature. |
WaterTemp |
false |
false |
Scalar |
READ |
Tango::DEV_FLOAT |
EXPERT |
Cryodrive water temp. |
EvapHeat |
false |
false |
Scalar |
READ |
Tango::DEV_UCHAR |
EXPERT |
Current evaporator heating. |
SuctHeat |
false |
false |
Scalar |
READ |
Tango::DEV_UCHAR |
EXPERT |
Current SUCT heating. |
RefrHeat |
false |
false |
Scalar |
READ |
Tango::DEV_UCHAR |
EXPERT |
Current refrigerator heating. |
ShieldHeat |
false |
false |
Scalar |
READ |
Tango::DEV_UCHAR |
EXPERT |
Current shield heating. |
NozzleHeat |
false |
false |
Scalar |
READ |
Tango::DEV_BOOLEAN |
EXPERT |
Nozzle heater value (on / off). |
AverageGasHeat |
false |
false |
Scalar |
READ |
Tango::DEV_UCHAR |
EXPERT |
Average value of gas heater. |
AverageSuctHeat |
false |
false |
Scalar |
READ |
Tango::DEV_UCHAR |
EXPERT |
Average value of SUCT heater. |
AverageNozzleHeat |
false |
false |
Scalar |
READ |
Tango::DEV_UCHAR |
EXPERT |
Average value of nozzle heater. |
EvapAdjust |
false |
false |
Scalar |
READ |
Tango::DEV_BOOLEAN |
EXPERT |
Evaporator adjusts vacuum compensation. |
TurboMode |
false |
false |
Scalar |
READ |
Tango::DEV_BOOLEAN |
EXPERT |
Turbo mode is active. The gas flow is increased to 10 l/min. |
ShutterState |
false |
false |
Scalar |
READ |
Tango::DEV_BOOLEAN |
OPERATOR |
Cryo shutter is closed. |
ShutterTime |
false |
false |
Scalar |
READ |
Tango::DEV_FLOAT |
OPERATOR |
Remaining anneal time. |
DewarLevel |
false |
false |
Scalar |
READ |
Tango::DEV_UCHAR |
EXPERT |
Current Dewar level. |
TimeToFill |
false |
false |
Scalar |
READ |
Tango::DEV_USHORT |
EXPERT |
Estimated time to the next Dewar refill. |
Vacuum |
false |
false |
Scalar |
READ |
Tango::DEV_FLOAT |
EXPERT |
The pressure measured by the vacuum gauge. |
VacuumGaugePower |
false |
false |
Scalar |
READ |
Tango::DEV_UCHAR |
EXPERT |
Power to vacuum gauge. |
HePressure |
false |
false |
Scalar |
READ |
Tango::DEV_FLOAT |
EXPERT |
Cryodrive He pressure |
SupplyPressure |
false |
false |
Scalar |
READ |
Tango::DEV_FLOAT |
EXPERT |
The pressure of the supplied gas. |
CryoStatus |
false |
false |
Scalar |
READ |
Tango::DEV_STRING |
EXPERT |
An indication of the status of the Cryodrive. |
CryoShift |
false |
false |
Scalar |
READ |
Tango::DEV_UCHAR |
EXPERT |
A parameter indicating the amount by which the Cryodrive speed has
been increased to allow very low temperatures to be attained. |
CryoSpeed |
false |
false |
Scalar |
READ |
Tango::DEV_UCHAR |
EXPERT |
The current Cryodrive speed. |
Suspend |
false |
false |
Scalar |
READ |
Tango::DEV_BOOLEAN |
EXPERT |
Flag to indicate device is in temporary Hold. |
OuterFlow |
false |
false |
Scalar |
READ |
Tango::DEV_DOUBLE |
EXPERT |
Current flow rate in outer stream. |
DetectedGasType |
false |
false |
Scalar |
READ |
Tango::DEV_STRING |
EXPERT |
Detected gas in cold stream. |
SelectedGasType |
false |
false |
Scalar |
READ |
Tango::DEV_STRING |
EXPERT |
Selected gas in coldstream. |
TotalHours |
false |
false |
Scalar |
READ |
Tango::DEV_USHORT |
EXPERT |
Total number of hours device has run. |