OxfordCryostream700 Class Attributes
Name
Inherited
Abstract
Attr. type
R/W type
Data type
Level
Description
GasSetPoint false false Scalar READ Tango::DEV_FLOAT OPERATOR Set temperature.
GasTemp false false Scalar READ Tango::DEV_FLOAT OPERATOR Current gas temperature.
GasError false false Scalar READ Tango::DEV_FLOAT EXPERT Difference between current and set temperature.
RunMode false false Scalar READ Tango::DEV_STRING EXPERT RunMode of the cryo controller (StartUp, StartUpFail, StartUpOk, Run, SetUp, ShutdownOk, ShutdownFail).
Phase false false Scalar READ Tango::DEV_STRING EXPERT Program Phase of the cryo controller (Ramp, Cool, Plat, Hold, End, Purge, Soak, Wait). This parameter is meaningless unless RunMode is Run.
RampRate false false Scalar READ Tango::DEV_USHORT OPERATOR Cooling ramp rate for the current program phase.
TargetTemp false false Scalar READ_WRITE Tango::DEV_FLOAT OPERATOR Temperature set point for the current program phase. If a value is written, the device will ramp up with 360K/h or cool down to it.
Remaining false false Scalar READ Tango::DEV_USHORT EXPERT Time remaining in current program phase.
GasFlow false false Scalar READ Tango::DEV_FLOAT OPERATOR Current gas flow.
GasHeat false false Scalar READ Tango::DEV_UCHAR EXPERT Current gas heating.
LinePressure false false Scalar READ Tango::DEV_FLOAT EXPERT Current back pressure.
Alarm false false Scalar READ Tango::DEV_STRING OPERATOR Indicator for most serious alarm conditions.
RunTime false false Scalar READ Tango::DEV_STRING EXPERT Uptime of the pump.
ControllerNumber false false Scalar READ Tango::DEV_USHORT EXPERT Serial number of the controller device.
SoftwareVersion false false Scalar READ Tango::DEV_UCHAR EXPERT Firmware version of the controller device.
Hardware false false Scalar READ Tango::DEV_STRING EXPERT Connected controller and cooler system devices.





OxfordCryostream700 Class Dynamic Attributes
Name
Inherited
Abstract
Attr. type
R/W type
Data type
Level
Description
ColdheadSerial false false Scalar READ Tango::DEV_USHORT EXPERT N-Helix coldhead serial number.
EvapTemp false false Scalar READ Tango::DEV_FLOAT EXPERT Current evaporator temperature.
SuctTemp false false Scalar READ Tango::DEV_FLOAT EXPERT Current SUCT temperature.
RefrTemp false false Scalar READ Tango::DEV_FLOAT EXPERT Current refrigerator temperature.
ShieldTemp false false Scalar READ Tango::DEV_DOUBLE EXPERT Current shield temperature.
NozzleTemp false false Scalar READ Tango::DEV_FLOAT EXPERT Current nozzle temperature.
WaterTemp false false Scalar READ Tango::DEV_FLOAT EXPERT Cryodrive water temp.
EvapHeat false false Scalar READ Tango::DEV_UCHAR EXPERT Current evaporator heating.
SuctHeat false false Scalar READ Tango::DEV_UCHAR EXPERT Current SUCT heating.
RefrHeat false false Scalar READ Tango::DEV_UCHAR EXPERT Current refrigerator heating.
ShieldHeat false false Scalar READ Tango::DEV_UCHAR EXPERT Current shield heating.
NozzleHeat false false Scalar READ Tango::DEV_BOOLEAN EXPERT Nozzle heater value (on / off).
AverageGasHeat false false Scalar READ Tango::DEV_UCHAR EXPERT Average value of gas heater.
AverageSuctHeat false false Scalar READ Tango::DEV_UCHAR EXPERT Average value of SUCT heater.
AverageNozzleHeat false false Scalar READ Tango::DEV_UCHAR EXPERT Average value of nozzle heater.
EvapAdjust false false Scalar READ Tango::DEV_BOOLEAN EXPERT Evaporator adjusts vacuum compensation.
TurboMode false false Scalar READ Tango::DEV_BOOLEAN EXPERT Turbo mode is active. The gas flow is increased to 10 l/min.
ShutterState false false Scalar READ Tango::DEV_BOOLEAN OPERATOR Cryo shutter is closed.
ShutterTime false false Scalar READ Tango::DEV_FLOAT OPERATOR Remaining anneal time.
DewarLevel false false Scalar READ Tango::DEV_UCHAR EXPERT Current Dewar level.
TimeToFill false false Scalar READ Tango::DEV_USHORT EXPERT Estimated time to the next Dewar refill.
Vacuum false false Scalar READ Tango::DEV_FLOAT EXPERT The pressure measured by the vacuum gauge.
VacuumGaugePower false false Scalar READ Tango::DEV_UCHAR EXPERT Power to vacuum gauge.
HePressure false false Scalar READ Tango::DEV_FLOAT EXPERT Cryodrive He pressure
SupplyPressure false false Scalar READ Tango::DEV_FLOAT EXPERT The pressure of the supplied gas.
CryoStatus false false Scalar READ Tango::DEV_STRING EXPERT An indication of the status of the Cryodrive.
CryoShift false false Scalar READ Tango::DEV_UCHAR EXPERT A parameter indicating the amount by which the Cryodrive speed has
been increased to allow very low temperatures to be attained.
CryoSpeed false false Scalar READ Tango::DEV_UCHAR EXPERT The current Cryodrive speed.
Suspend false false Scalar READ Tango::DEV_BOOLEAN EXPERT Flag to indicate device is in temporary Hold.
OuterFlow false false Scalar READ Tango::DEV_DOUBLE EXPERT Current flow rate in outer stream.
DetectedGasType false false Scalar READ Tango::DEV_STRING EXPERT Detected gas in cold stream.
SelectedGasType false false Scalar READ Tango::DEV_STRING EXPERT Selected gas in coldstream.
TotalHours false false Scalar READ Tango::DEV_USHORT EXPERT Total number of hours device has run.