Talks to the ISEG HV power supply.
e.g.:
* = vhq205l( hv1, set_point)
* = vhq205l( hv1, set_point, 123)
Here is the list of keywords. 'RW' means that a keyword can be read and written.
Keywords:
actual_current, R
actual_value, R
current_limit, muA, RW
data_ready, R
D0 voltage A
D1 current A
D2 voltage B
D3 current B
hw_limits, switches from the fron panel, R
D0..D3 current
D4..D8 current
modul, R
ramp, V/s, [2,255], RW
register_1, R
D0 Output voltage channel A is zero
D1 Control switch, 0 -> DAC, 1 -> manual
D2 Output polarity, 1 -> positive
D3 HV on (0)/off (1)
D4 Kill switch, 0 -> disabled
D5 Output voltage derivative, 1 -> rising
D6 Status V-out, 1 -> V-out changing
D7 Error on channel A, 1 -> error
D8 ... channel B
register_2, R
D0 Timeout error
D1 I-out > I-max
D2 V-out reached set point
D3 A front panel switch has been changed
D4 Ratio V-setpoint/V-max > 1
D5 External INIHIBT is/was active
D6 Exceed/ed V-max or I-max
D7 Quality of V-out not garanteed
D8 ... channel B
set_point, RW