pogo_html_docu/I404/Attributes.html See pogo generated info for more details.
Position_x The position in the PETRA III coordinate system. |
Position_y The position in the PETRA III coordinate system. |
Factor_x position_x = read_position_x |
Factor_y position_y = read_position_y |
Calib_source Set/query internal calibration source state, 0 - off, 1 - to channel 1, ..., 4 - to channel 4. |
Conf_capacitor Selects the capacitor: 0 - 100 pF, 1 - 3300 pF. Automatically changed when a new range is set. |
Conf_hv_max Set/query maximum allowable hv setting in volts. |
Conf_hv_set Set/query the external hv in volts. |
Conf_intavg Sets the no. of integrations per reading, 1 .. 15. |
Conf_monitor Set/query the position calculation function. 1 no position calculation, 2 - quadrant mode calculations, 3 - split mode calculations |
Conf_period Set/query the integration period in seconds. Automatically changed when a new range is set. |
Conf_pos_polarity Set/query the polarity of the current readings used for the calculations 0 - +ve, 1 - -ve |
Conf_pos_threshold Set/query (in per cent of full current scale in use) for data to be included in position calculations. |
Conf_range Set/query the full scale current range in amps, integration period and capacitor selection are calculated by the I404. |
Conf_readavg Set/query the number of adc readings to be taken in each integration period. |
Conf_resolution Set/query the number of bits of effective resolution, 16 - 20, the i404 calculates the number of integration periods and adc readings per integration. |
Read_currentI Query current reading of channel I, I = 1 .. 4. The currents are updated once per second. |
Read_integrationtime Query the used integration time. |
Read_hv Query hv output sense. |
Read_position_x Query the x-position (first value) which was calculated in the configured mode. |
Read_position_y Query the y-position (second value) which was calculated in the configured mode. |
Syst_frequency Set/query the dominant noise frequency in Hz to be suppressed in the calibration routine, 50 or 60 Hz |
Syst_save Set/query save state flag, 0 do not go to safe state when disconnected, 1 - go to safe state when disconnected (hv off) |
Syst_serial Query the systen serial number.
|